14
DO NOT force the handle beyond the upper or lower stops. The hydraulic unit can be
rotated to allow the handle to be used from either side of the lift. To lower the boom,
turn the release valve anticlockwise. The release valve is progressive, the more it is
opened the faster the descent. The valve is restricted so even when fully open the
descent is controlled. This facility allows for a "hands free" descent. If the release
valve is opened a fraction (a quarter turn) a very slow speed of descent will allow the
caregiver to work "hands free" while assisting or comforting the patient. REMEMBER to
close the release valve before commencing lifting operations. The release valve only
requires gentle pressure to open or close. DO NOT apply excessive force to the release
valve, either to close or to open. It is not necessary and will only damage the valve.
9. Slings
The selected sling is attached to the spreader bar hooks, each sling is supplied with
instructions. Please study the instruction guide for the lifter and sling before using. Once
the correct sling has been selected for the patient, attach it to the spreader bar hooks.
CHARGING INSTRUCTIONS FOR THE OXFORD/HOYER ADVANCE LIFT
On Board Charging Option
The batteries are located in the power pack and are charged through a socket at the
base of the control box.
1. Insert the power supply plug into the charge point at the base of the control box.
The plug is inserted with a straight push. DO NOT twist the plug in the socket.
2. Plug the charger mains lead/cord into a suitable mains/power outlet and switch
the mains/power supply ON.
3. Charging is fully automatic. The LCD indicator indicates the status of the charg-
ing. Note: Even if the charger is left plugged in for extended periods it will not
allow the batteries to "overcharge".
4. To return the lift to service, switch OFF the mains supply, pull the charger socket
out and disengage from the mains supply
294000.10003 Rev. B
E n g l i s h
Removable Battery Pack
Charge Point
E n g l i s h
Off Board Charging Option
There is an off board charging option on the Advance hoist. This can be activated by
inserting the relevant charge lead into the controller base and connecting this lead to
the mains supply. Charging is then automatic. To return the lift back to service simply
remove the lead. The battery indicator will then highlight the amount of charge that
has been provided to the batteries. It is advised to charge the batteries fully before
returning the lift to service.
• Do NOT use the lift whilst charging is taking place.
• Be careful not to trip over the charge lead.
WARNING check the hoist is not charging before moving as the electrical
•
connection may be damaged
CAUTION: Keep the batteries fully charged. Place the lift on charge whenever
•
it is not in use. The charger will not allow the batteries to "overcharge".
CAUTION: Never run the batteries completely flat. If the audible warning
•
sounds, complete the lifting operation in hand and place the lift on charge.
CAUTION: Never store the lift for long periods without regular charging
•
throughout the storage period.
CAUTION: Always make sure the mains power to the charger is switched off
•
before connecting or disconnecting the charger to or from the lift.
CAUTION: Never leave the charger plugged into the lift with the mains
•
power off.
CAUTION: Never disconnect the charger plug by pulling on the cable.
•
• Do not charge an electric lift in a bathroom or shower room.
WARNING
294000.10003 Rev. B
15