Aesculap
ELAN EC motor with right/left rotation (without irrigation unit/with irrigation
unit)
Validated cleaning and disinfection procedure
Validated procedure
Special features
■
Wipe disinfection for
electrical devices with-
out sterilization
5.6
Wipe disinfection for electrical devices without sterilization
Phase
Step
I
Wipe disinfection
RT:
Room temperature
Phase I
►
Remove any visible residues with a disposable disinfectant wipe.
►
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
►
Observe the specified application time (1 min minimum).
5.7
Inspection, maintenance and checks
►
Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged.
►
Set aside the product if it is damaged.
6.
Maintenance
To ensure reliable operation, the product must be maintained at least once
a year, including the power cord and foot control.
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
12
®
Keep the coolant pump open during cleaning and close again after
cleaning.
T
t
Conc.
[°C/°F]
[min]
[%]
≥1
RT
-
Water quality
Chemical
-
Meliseptol HBV wipes 50 % Propan-1-ol
Reference
Chapter Wipe disinfection for electrical devices
without sterilization