Technical Information
Profibus Message Mapping
Offset
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
344/360
Meaning
Reserved
Master Add (master address after parameterisation. FF= without
parameter)
Ident number high
Ident number low
External diagnosis "Header" length indication including header
Interface Status
B
Controller not responds
0
B
Controller not identified
1
B
Input data not fully coherent yet
2
B
Not used
3
B
Not used
4 - 7
B
Start/Stop diag status
7 - 4
B
Low Speed diag status
3 - 0
B
Soft Start diag status
7 - 4
B
Active Stop diag status
3 - 0
B
Vent Valve diag status
7 - 4
B
Purge diag status
3 - 0
B
Heater Jacket diag status
7 - 4
B
Stop Speed Reading diag status
3 - 0
B
High speed setting diag status
7 - 4
B
Low speed setting diag status
3 - 0
B
Water Cooling Max Temp diag status
7 - 4
B
Water Cooling Hyst diag status
3 - 0
B
Set Point Hyst diag status
7 - 4
B
Heater Jacket Setting diag status
3 - 0
B
Flow Meter Threshold diag status
7 - 4
B
Reserved diag status
3 - 0
B
Set Point Value diag status
7 - 4
B
Set Point Delay diag status
3 - 0
B
Vent Valve Open Delay diag status
7 - 4
B
Vent Open Time diag status
3 - 0
B
Vent/Purge Auto/Man diag status
7 - 4
B
Vent/Purge Remote/Profibus diag status
3 - 0
Turbo-V 2K-G Pumping System User Manual / 87-900-968-01 (H)