16
Instructions for Use
Installation
FLANGE
ISO 250 F
The Turbo-V 2K-G cannot be fixed by means of its base. The system can only be
NOTE
fixed through its ISO 250F flange. Fixing must be done according to ISO 1609
norm. Steel bolts with a strength class at least of 8.8" must be used.
The Turbo-V 2K-G belongs to the second installation (or overvoltage) category as
CAUTION!
per directive EN 61010-1. Connect the device to a mains line that satisfy the
above category. The Turbo-V 2K-G has Input/Output and serial communica-tion
connectors that must be connected to external circuits in such a way that no
electrical part is accessible. Use the provided mating connectors in order to
maintain the IP-54 protection level. Be sure that the insulation of the device
connected to the Turbo-V 2K-G is adequate even in the case of single fault as per
directive EN 61010-1.
For installation of optional accessories, see "Technical Information".
290/360
FIXING DEVICE
M10 bolts
Turbo-V 2K-G Pumping System User Manual / 87-900-968-01 (H)
N.
FIXING TORQUE
12
22 Nm