Turbo-V 2K-G Pumping System User Manual / 87-900-968-01
Range
Description
B0 - Vent & Purge Auto/Man
0 = Auto (the controller drives purge and vent valves)
1 = Manual (see Par. Buff. byte-15-B2)
B1 - Vent & Purge Remote/Profibus
0 = Serial or Profibus (see Out. Buff. byte-0 B3 and B2)
1 = Remote (the valves are driven by the inputs on
remote I/O connector)
B2 - Interlock Type
0 = continuos (the interlock must be present to run the
pump)
1 = at start (the interlock must be present only at start
command)
B3 - Setpoint Logic
0 = active close (NO)
1 = active open (NC)
B4 - Heather Jacket Mode (Auto/man)
Specific option, not available on standard models
B5 - Gas Load Type
0 = Ar, 1 = N
2
B6 - Purge Gas Type
0 = N
1 = Ar
2,
B7 - Reserved -
B0. B1 - Reserved -
B2. B3 - Start/Stop/Interlock Mode
Change the logic of Start/stop and Interlock inputs.
Ask to Agilent personnel for this option
B4.. B5 - Set Point Type
0 = frequency
1 = current
2 = time
3 = status (normal operation)
B6. B7 - Analog Output Type
Set the analog output on remote I/O connector
proportional to:
0 = frequency
1 = power