N.
Read/ Write
147
R/W
152
R/W
155
R
157
R/W
158
R/W
159
R/W
164
R
200
R
201
R
202
R
203
R
204
R
205
R
206
R
211
R
216
R
Turbo-V 2K-G Pumping System User Manual / 87-900-968-01
Data
Description
Type
N
Vent Valve Opening Time
(expressed in 0.2 sec)
L
HPS enable
N
Power limit applied
L
Gas load type
N
Max foreseen water cooling
temperature (Twater,max)
N
Water cooling hysteresYs
N
Tbody,lim: body temperature corresponding to Twater,max
N
Pump current in mA dc
N
Pump voltage in Vdc
N
Pump power in W (pump current
x pump voltage duty cycle)
N
Driving frequency in Hz
N
Pump temperature in °C
N
Pump status
N
Error code
N
Controller temperature n. 1
N
Pump body temperature
Technical Information
Serial Cable Installation
Admitted
Values
0 to 65535 bit
(0=infinite, 1=0.2sec,
max=13107 sec)
(default =0)
0 = OFF
1 = ON
(default = 0)
write only in STOP
Watt
0 = Ar
1 = N
2
(default = 0)
°C, min = 15
max = 35
(default = 34)
% (default = 1)
0 to 70
Stop = 0
Waiting intlk = 1
Starting = 2
Auto-tuning = 3
Braking = 4
Normal = 5
Fail = 6
Bit description: see
the following figure
°C
°C
(H)
17