17
Technical Information
Serial Cable Installation
N.
Read/ Write
111
R/W
114
R/W
115
R/W
117
R/W
118
R/W
120
R/W
121
R/W
122
R/W
123
Reserved to Agilent service
124
125
R/W
126
R/W
130
Reserved to Agilent service
145
R/W
146
R/W
326/360
Data
Description
Type
L
Analog output type: output
voltage signal proportional to
frequency or power
N
Purge gas type
L
Threshold flowmeter
N
Low speed setting (Hz)
L
Body thermistor enable
N
High speed setting (Hz)
N
Maximum excitation frequency
in Hz (active only in Stop
condition)
L
Set vent valve open/close (N.C.)
L
Valves operating mode
N
Vent valve opening delay
(expressed in 0.2 sec)
L
Set purge valve open/close
L
Set valves command source
Turbo-V 2K-G Pumping System User Manual / 87-900-968-01 (H)
Admitted
Values
0 = frequency
1 = power
(default = 0)
0 = N
2
1 = Ar
sccm
475 to win. 121
(default = 475)
0 = NO
1 = YES
475 to win. 121
(default = 555)
475 to 555
(default = 555)
1 = open
0 = close
(default = 0)
0 = auto (default)
1 = on command
(see win.146)
0 to 65535
(corresponding to 0
to 13107 sec)
(default = 15)
0 = close
1 = open
(default = 0)
0 = Serial
1 = Remote
(default = 0)