17
Technical Information
Profibus Message Mapping
22
1
°C
23
1
°C
24
1
°C
25
1
°C
26
1
-
27
1
Bit
Field
340/360
0-45
Pump Body Temperature
Read the pump body temperature close to the water circuit.
If > "Pump Body Temperature Limit" (In Buff byte-25) the
controller will reduce the "Max Allowable Power"
If > 38°C the controller goes in fail "pump over-temperature"
and the pump is stopped
0-100
Controller Power Section Temperature
Read the controller power section temperature if > 75°C the
controller goes in fail "controller over-temperature"
0-100
Controller CPU Section Temperature
Read the controller CPU section temperature if > 80°C the
controller goes in fail "controller over-temperature"
15-38
Pump Body Temperature Limit
Read the body temperature limit calculated from the" Water
Cooling Max Temperature" (parameter buffer byte-0)
0-6
B0-B3 – Status
0 = Stop
1= Waiting for interlock (interlock connections on remote
I/O connector missing)
2 = Ramp (the pump is increasing the speed)
3 = Auto tuning (the speed has been reduced because the
gas load is higher than the Power Limit)
4 = Braking (the pump is reducing the speed using the
motor)
5 = Normal operation (the pump is at target speed)
6 = Fail (see error code byte Input Buffer Byte-27)
B4-B7 – Reserved
Error Code (bit field)
B7 – Too High Load (the gas load is too high for the pump)
B6 – Short Circuit (a short circuit happened between the
two of the three motor phases)
B5 – Over-voltage (not used)
B4 – Verify Line Select (the auto voltage selection
mismatch the real input voltage)
B3 – Power Fail (internal circuitry failure)