Window Meanings
Tab. 8
N.
Read/ Write
000
R/W
008
R/W
100
R/W
101
R/W
102
R/W
103
R/W
104
R/W
105
R/W
108
R/W
109
W
Turbo-V 2K-G Pumping System User Manual / 87-900-968-01
Data
Description
Type
L
Start/Stop
(in remote mode the window is a
read only)
L
Remote
(default) or Serial configuration
L
Soft Start
(write only in Stop condition)
N
Set Point type
N
Set Point threshold
(expressed in Hz, mA or s)
N
Set Point mask: time between
the pump start and the set point
check (seconds)
L
Set Point signal activation type:
the signal can be "N.O." or
"N.C."
N
Set point hysteresis
(in % of threshold)
N
Baud rate
L
Pump life/ cycle time/ cycle
number reset
Technical Information
Serial Cable Installation
Admitted
Values
Start = 1
Stop = 0
Remote = 1
Serial = 0
(default = 1)
YES = 1
NO = 0
0 = Frequency
1 = Current
2 = Time
(default = 0)
(default = 500)
0 to 99999
(default = 0)
0 = active close
(N.O.)
1 = active open (N.C.)
(default = 0)
0 to 100
(default = 2)
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
To reset write '1'
(H)
17
325/360