Technical Information

Profibus Message Mapping

Profibus Message Mapping
Tab. 12
Parameter buffer (17 bytes)
Offset
Size
Unit Res
Byte
Byte
0
1
°C
1
1
°C
2
1
%
3
1
°C
4
1
Sccm
5
2
-
7
2
Hz/mA/s
9
2
s
11
2
0.2*s
13
2
0.2*s
336/360
Range
Description
15-35
Water Cooling Max Temperature
Enter here the max temperature you foreseen for your
cooling system, this value will set the maximum power
provided to the pump (see Input Buffer byte-4)
1-5
Water Cooling Hysteresis
Set here the hysteresys on water cooling measurement
(default=1, don't change if not necessary)
1-99
Setpoint Hysteresis
Set the hysteresis on set point output (see set point out
paragraph)
-
Heather Jacket Temperature Setting
Specific option, not available on standard models
-
Flowmeter Threshold
Specific option, not available on standard models
-
Reserved
-
Setpoint Value
Set the threshold for the set point out
0-
Setpoint Mask Time
99999
Set the time during the controller does not drive the set
point out
0-
Vent Valve Opening Delay
65535
Set the delay time between the stop command and vent
0 = inf.
opening. Valid only in "purge/vent auto mode"
(Parameter Buffer byte-15 B1=0); default = 15;
0 = infinite (the valve will never open)
0-
Vent Valve Open Time
65535
Set how long the vent will remain open, then both purge
0 = inf.
and vent will be closed. Valid only in "purge/vent auto
mode" (Parameter Buffer byte-15 B1=0); 0 = infinite
(the valve remain open till a start command)