Agilent Technologies Turbo-V 2K-G Notice De Mode D'emploi page 341

Publicité

28
1
Sccm
29
1
-
Turbo-V 2K-G Pumping System User Manual / 87-900-968-01
B2 – Controller Over-Temperature
("Controller Power Section Temperature" or "Controller
CPU Section Temperature" are over the limit)
B1 - Pump Over-temperature
("Pump Bearing Temperature" or "Pump Body
Temperature" are over the limit)
B0 – Check Connection To Pump
(the connection between controller and pump missing)
Flow Meter Reading
(Specific option, not available on standard models)
B0 – Start/Stop Status
0 = Stop
1 = Start
B1 – Low Speed Status (Read the "Low Speed" function
status (Out Buff byte-0 B6))
0 = off (the "target speed" (In Buff byte-8) is" high speed"
(Out. Buff. byte-1))
1 = on (the "target speed" (In Buff byte-8) is "low speed"
(Out. Buff. byte-3))
B2 – Soft Start Status
0 = on
1 = off
(this function is active at first pump run-up and can't be
deactivated by the user, after first run-up activate this
function only if the pump has been stopped for at least3
months; this procedure will take 1h)
B3 – Vent Valve Status
0 = close, 1 = open
B4 – Purge Valve Status
0 = close, 1 = open
B5 – Set Point Status
0 = open, 1 = closed
B6 – Flow meter Alarm Status
Specific option, not available on standard models
B7 – Not Used
Technical Information
Profibus Message Mapping
(H)
17
223
000
001
100
122
145
221
212
341/360