Medela Thopaz+ Mode D'emploi page 12

Masquer les pouces Voir aussi pour Thopaz+:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 103
Battery low
Approximately 30 minutes before the battery is fully empty an acoustic signal sounds and the
battery symbol [
] starts to blink. The acoustic signal can be muted but the battery symbol
continues to blink until the battery is empty. The set pressure is maintained but the battery should
be recharged as soon as possible.
Battery empty
The battery symbol will blink with an acoustical sound for 10 minutes prior to the battery fully
discharging, unless the pump is switched off earlier. The acoustic sound cannot be muted during
the 10 minutes. If Thopaz + is not connected to a power source the pump will switch off automati-
cally after 10 minutes. The negative pressure is not maintained after the pump turns off.
*
1. Battery empty
*
2. Battery almost empty
3. Battery is being charged (moving bars)
4. Battery is fully charged and connected to the mains
* Symbol blinks
12 |
Power supply and battery operation
Press simultaneously to acknowledge an acoustic alarm and
follow instructions on the display (see chapter 12 – Trouble-
shooting and alarms).