Visible on display
Error
Problem
number
description
306
Canister full
401
Battery low
402
USB connection
not permitted
405
Standby
406
Temperature too
high
409
Fluid level
sensor out
of order
410
Canister
detection
301
Leak in system
302
Tubing clogged
305
Battery empty
311
Selftest failed
313
Filter clogged
Thopaz +
315
overheated
407
Fluid alarm
Troubleshooting
instructions
1. Clamp tubing
2. Switch to Standby
3. Replace canister
4. Continue with "On"
5. Unclamp tubing
Connect Thopaz + to mains
Unplug USB cable
Switch Thopaz + on or off
Prepare backup device
– Fluid measurement
unavailable
– Confirm and continue
therapy
1. Clamp tubing
2. Switch to Standby
3. Replace canister
4. Continue with "On"
5. Unclamp tubing
Check system for leaks
1. Clamp catheter
2. Clamp tubing
3. Switch to Standby
4. Replace tubing
5. Continue with "On"
6. Unclamp catheter
Connect Thopaz + to mains
Snap the canister out and
in again
1. Clamp tubing
2. Replace canister
3. Continue with "On"
4. Unclamp tubing
Replace Thopaz +
Patient is losing a large
amount of fluid
Remarks / Potential source of
error
Remaining time of battery
approx. 30 min.
USB connection not
permitted during operation
or in the vicinity of the
patient.
In Standby for 5 minutes.
Do not expose Thopaz + to
heat sources.
– Disconnection.
– Missing seal at vacuum
port.
– Tubing is kinked or
clogged for approx.
10min.
– Thopaz + detects clotting
of Thopaz + tubing, NOT
of the patient catheter.
Do not switch Thopaz + on if
the patient is already
connected.
Filter in canister clogged.
Troubleshooting and alarms
Pressure
Yes
Yes
Yes
No
Yes
Yes
Yes
No
No
No
No
No
No
Yes
| 43