9 Changing pressure during operation
WARNINGS
The negative pressure may only be changed by the physician or on medical orders.
Change pressure
9.1
Physiological
pressure
38 |
Changing pressure during operation
Important: Thopaz + is running.
Press simultaneously
For patients who are to be treated by gravity drainage (= water
seal), the physiological pressure can be activated.
1.1 Press [
] and [
simultaneously.
1.2 Change pressure by
pressing [
] or [
and confirm with "OK"
[
].
1.1 Press "Physio" [
1.2 Confirm with "OK" [
This mode corresponds to a
pressure of -0.8kPa/
-6 mmHg/-8cmH2O/-8 mbar.
]
]
].
].