Visible on display
Error
Problem
number
description
Internal error
XXX
Safety related tip
The acoustic alarm "Leak in system – 301" is deactivated for approximately 6 minutes
after Thopaz + is switched on.
Safety related tip
Warning "canister full"/Alarm "filter clogged".
The warning "canister full" is triggered when the canister filling level approximately reaches the
maximal scale line indicated on the respective canister label. Change the canister according the
instructions of use, chapter 8 – Replace canister. If secretions line the walls of the canister, the
warning "canister full" may be activated prematurely.
After muting the warning with the selection buttons, a symbol [
After 5 minutes the warning reappears acoustically.
An overfilling of the respective canister volume will trigger the alarm "filter clogged". The alarm
is triggered due to the clogging of the hydrophilic filter upon contact with liquids or secretions.
Please refer to chapter "4 – Overview" for location of the hydrophilic, overflow protection/
bacteria filter in the canisters. Subsequent to a clogged filter and the alarm "filter clogged"
the vacuum applied to the patient is interrupted and cannot be regulated or maintained.
An additional accumulation of liquids, secretions and air is released through the pressure relieve
valve at the positive pressure of 0.2 - 0.5 kPa. This pressure release function is not a substitute for
a free overflow function. Therefore the canister must be replaced once full. Please refer to
chapter "4 – Overview" for location of the pressure relieve valve.
Change the canister according to the instructions for use, chapter "8 – Replace canister".
44 |
Troubleshooting and alarms
Troubleshooting
instructions
1. Switch Thopaz + off and on
again
2. Inform Medela Customer
Service
Remarks / Potential source of
error
If error message repeats,
inform Medela Customer
Service.
] appears during operation.
Pressure