4 OPERATION

4.2 OPERATION MODE DISPLAY

The following variables are displayed in the operation mode:
45,6 L/m
16,45 m A
0......9
80529 L
6247 L . .

4.3 CALIBRATION MODE

Press
simultaneously for 5 seconds
ENTER
The following variables can be set in the parameter definition menu:
LANGUAGE
UNIT
K-FAKTOR
CURRENT
0......9
PULSE
RELAY
FILTER
TOTAL
END
Flow rate in the required engineering unit (see calibration menu).
4-20 mA output signal, proportional to the flow
according to the selected measuring range
Main totalizer in the required engineering unit (see calibration
menu). Reset in the calibration menu.
Daily totalizer in the same engineering unit as the main totalizer. A
point behind the unit differentiates it from the main totalizer. Reset
by simultaneously pressing the
Language selection between English, German, French and Italian.
Selection of engineering units to display flow rate and totalizer.
Input of K-factor according to chart or Teach-in function in order
to determine the specific K-factor.
Setting of the 4-20 mA measuring range.
Parameter definition of pulse output (unit and quantity).
Parameter definition of relays. This message only appears if the
relay option has been installed.
Dampening selection. There are ten different steps and 2 filtering
modes available.
Selection of the min. input frequency.
Totalizer resetting.
Return to operation mode and storage of new parameters.
E-22-
8025 UNIVERSAL