4 OPERATION
4.3.8 Totalizer
The main and daily totalizers are reset in this menu. The reset procedure only starts when
Enter is pressed, at the "END" position in the parameter definition menu.
TOTAL
ENTER
END
ENTER
4.4 TEST MENU: PRESS
The following compensations and tests are carried through in the test menu:
OFFSET
SPAN
0......9
FREQUENC
FLOW
END
4.4.1 Offset-compensation
Here, the customer has the option to correct the basic setting of 4 mA. He only needs one
current meter. When Enter is pressed while "OFFSET" is displayed, the transmitter produces
4 mA. If this value is incorrect, it can be corrected by entering the measured value.
OFFSET
ENTER
SPAN
ENTER
RES NO
RES YES
ENTER
0......9
Zero-compensation (4 mA).
Span-compensation (20 mA).
Indication of sensor frequency.
Entering the flow rate to be simulated. The outputs will react in
accordance to this input.
Return to the operation mode and storage of new parameters for
OFFSET and SPAN. If one of the two values is erroneous, the
device point on "OFFSET", and new values must be entered.
OF = 4.00
0......9
OF = 4.02
Entering the measured value
E-28-
8025 UNIVERSAL
SIMULTANEOUSLY FOR 5 S.