Option A13
RS485
Connector 69Y03166
If necessary, a terminating resistor 470R, 1%, 0,6W
may be soldered between Pin J and K of the connector
2)
CIXS3
Mainboard
der CIXS3
RxD-TxD-P
RxD-TxD-N
Signal_GND
LV4
RxD-TxD-P
2
RxD-TxD-N
3
1
Signal_GND
Important note!
Max. voltage V
of any electrical apparatus connected to the interface converter type YDI05-Z.. may not exceed 250V. The internal
m
Zener voltage V
is 12.6V.
z
Note:
1:
not in the scope of delivery
2: up to 8 devices; see „Verification of Intrinsic safety" (in the manual of these devices)
3: A cable can be rooted diretly into the housing via cable gland YAS04CIS instead of using the 12-contact connector (Option M58).
2013-02-02
Dr. D. Klausgrete
Hazardous area
14-contact female
connector on
SIWXS.../ISX...
Junction Box
1)
Cable gland
Junction Box
72583
connector
J
K
M
cable gland
RS485 connection with YDI05-Z
Title
Drawing number
Non-hazardous area
e.g.
72583
1)
screened cable
(e.g. YCC485-X)
up to 1000m
e.g.
72583
YDI05-Z
1
1
LV2
4
4
LV1
5
1
LV3
Data Transfer
36953-751-40
Options de sortie de données
YDI05-Z
LV3
Options
A24, M50,
3)
M55
LV2
LV1
4
1
1
6
E
C
B
12-contact female
connector on YDI05-Z..
RxD-TxD-P
C
B
RxD-TxD-N
E
Signal_GND
5
6
Page
of
00
Revision
Signum 3 Ex
135

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

SiwxsbbsSiwxsdcpSiwxsdcs

Table des Matières