Minebea Intec Signum 3 Ex SIWXSBBP Mode D'emploi page 117

Balances complètes pour les atmosphères explosibles
Table des Matières

Publicité

Canadian Certificate Of Conformity No: FM17CA0166X
9.
This certificate relates to the design, examination and testing of the products specified herein. The FM Approvals
surveillance audit program has further determined that the manufacturing processes and quality control
procedures in place are satisfactory to manufacture the product as examined, tested and Approved.
10.
Equipment Ratings:
SIWXSabc-d-ef-ghi. Weighing Unit.
SIWXSabc-d-Vklm. Weighing Unit.
ISXabc-ef-ghi. Weighing Unit.
ISXabc-Vklm. Weighing Unit.
IP6X Models: SIWXS Scale and ISX Weighing platform as Intrinsically Safe for use in Class I, II and III,
Division 1, Groups A, B, C, D, E, F and G; Class I, Zone 1, Ex ib Group IIC in accordance with Control
Drawing 36953-751-07; Temperature Class T4; Hazardous Indoor Locations over a temperature range of -
10°C to +40°C.
SIWXSBBP-3-06-Hhi. Weighing Unit.
SIWXSBBP-3-V1lm. Weighing Unit.
ISXBBP-3-06-Hhi. Weighing Unit.
ISXabc-V1lm. Weighing Unit.
IP4X Models: SIWXSBBP-3-06-H & SIWXSBBP-3-V1lm Scale and ISXBBP-3-06-H & ISXabc-V1lm Weighing
platform as Intrinsically Safe for use in Class I, Division 1, Groups A, B, C and D; Class I, Zone 1, Ex ib Group
IIC; in accordance with Control Drawing 36953-751-07; Temperature Class T4; Hazardous Indoor Locations
over a temperature range of -10°C to +40°C.
11.
The marking of the equipment shall include:
SIWXSabc-d-ef-ghi. Weighing Unit.
SIWXSabc-d-Vklm. Weighing Unit.
ISXabc-ef-ghi. Weighing Unit.
ISXabc-Vklm. Weighing Unit.
IS CL, I, II, III, DIV 1, GP A,B,C,D,E,F,G T4
CL I, Zone 1, Ex ib IIC T4
-10°C < Ta < +40°C
SIWXSBBP-3-06-Hhi. Weighing Unit.
SIWXSBBP-3-V1lm. Weighing Unit.
ISXBBP-3-06-Hhi. Weighing Unit.
ISXabc-V1lm. Weighing Unit.
IS CL, I, DIV 1, GP A,B,C,D T4
CL I, Zone 1, Ex ib IIC T4
-10°C < Ta < +40°C
12.
Description of Equipment:
General - The SIWXS scale series consists of a weighing module (platform) and a display unit located in the
hazardous area. The platform series ISX consists of a weighing module only and is located in the hazardous area.
Construction - The housing of the display unit consists of stainless steel with a membrane switch adhered to the
front panel. The display housing is two pieces and uses a gasket to prevent ingress of dust for IP6X model types.
The weighing modules consist of painted aluminium and/or stainless steel.
THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE
FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA
T: +1 (1) 781 762 4300 F: +1 (1) 781 762 9375 E-mail:
F 348 (Mar 16)
SCHEDULE
information@fmapprovals.com www.fmapprovals.com
FM/CSA Certificat
Page 2 of 3
Signum 3 Ex