Inserire a pag. 51 capitolo 5.1.9
5.1.9 Lubricating and cleaning the screw and nut screw
5.1.9 Lubrificazione e pulizia vite e boccola filettata
Every 50 cycles, clean
and lubricate with silicone
grease the inside of the
threaded bushing "A"
located on the door .
EN
Clean screw "B" on the front panel using
a soft cloth.
Never use hydrochloric acid or detergents containing
chlorine to clean any part of the autoclave.
5.1.10 Clean the feed tank
This must be carried out by qualified technicians.
5.3 Periodic process effectiveness and safety checks
Before proceeding with the operations indicated above, always switch off the main switch (5 Fig. 1) and make sure that the
chamber is cold.
Never use hydrochloric acid or detergents containing chlorine to clean any part of the autoclave.
70
Inserire a pag. 51 capitolo 5.1.9
5.1.9 Lubrificazione e pulizia vite e boccola filettata
A
                                                   
B
Operation
Vacuum Test EN 17665
Bowie & Dick Test 17665
Helix Test EN 17665
Protective earthing continuity
Insulation test
Casing leakage current
Initial validation EN 17665
Periodic validationa EN 17665
Safety valve check
5.2 Maintenance
After every 1000 cycles or two years, a reminder message appears
to perform maintenance.
The maintenance must only be carried out by an
authorised and qualified service technician.
The maintenance consists of replacing, cleaning and
checking some parts, as indicated in the relative table.
The work consists of the following:
Replace:
- door seal
- bacterial or Bio X filter
- chamber filter
- water filter
Check:
- Safety valve check
- Leakage current check
- Hose checks: hoses performance does not change until they
                                                   
physically break, in which
case they do not work. Only replace them with DÜRR DENTAL
spare parts.
When finished, the qualified technician will issue a certificate stating
that the autoclave maintenance was carried out.
Comply with the intervals specified or indicated in this manual and
in the additional sheets. Only use ORIGINAL SPARE PARTS. Failure
to comply will void the equipment warranty.
Frequency
Before
Daily
Annually
first use
Every two
years