4.7.1 Test program types
Selecting the "test" sub-menu accesses the list of test cycles that
autoclave can run:
- Vacuum Test
- 134°C Helix Test
- 121°C Helix Test
- 134°C Bowie&Dick Test
4.7.2 Vacuum test (leakage test)
To run this function, the chamber must be cold and have no instru-
ments in it.
4.7.3 Bowie & dick test
To run this test, the chamber must have no instruments in it. The
EN
Bowie & Dick "package" used for the porous load test must be
positioned on the lower tray.
4.7.4 Helix test
To run this test, the chamber must have no instruments in it. The
instrument used to test narrow hollow loads must be positioned on
the lower tray.
4.7.5
Diagrammi test
4.7.5 Test diagrams
Hygoclave 40/Hygoclave 40 Plus
Diagram 134°C Helix e Bowie & Dick Test
Pressure
Time
Hygoclave 40/Hygoclave 40 Plus
Diagram 121°C Helix Test
39 39
Pressure
Time
Hygoclave 40/Hygoclave 40 Plus
Diagram Vuoto Test
38
39 39
Pressure
Time
60
39 39
39 39