Fault
is displayed.
is displayed.
bAtt
StOP
is displayed.
Er:[No.]:11
is displayed.
is displayed.
Er:[No.]:12
is displayed.
Er:[No.]:16
Er:[No.]:32
is displayed.
9.2
Problems with data transmission (interface module)
Fault
Workflow LED does not
come on
The Workflow LED illumi-
nates in red, the measuring
result is displayed
The Workflow LED illumi-
nates in red, an error code
(
Er:8:[No.]
) is displayed
is displayed
Er:8:94
is displayed
Er:8:95
Er:[No.]:71
is displayed after
the
key is pressed
44 • seca 787/797
Cause
Battery voltage is dropping.
The batteries have discharged.
Maximum capacity has been exceeded.
The device has too high a load or too
high a load in one corner.
The device has been switched on with
too high a load.
The device was caused to oscillate, the
zero point could not be determined.
An internal communication error oc-
curred.
NOTE
Observe the system instructions for use of
structions for use of the barcode scanner.
Cause
The WiFi function is deactivated
Workflow LED is defective
Errors occurred during data transmis-
sion
Errors occurred during the measure-
ment procedure or data transmission
An error occurred during the measure-
ment procedure: IDs were scanned at
the wrong time
Hardware fault
The WiFi function is deactivated
Remedy
• Replace the batteries.
• Replace the batteries.
• Remove the load from the device.
• Remove the load from the device or dis-
tribute the weight more evenly.
• Switch off the device, then switch it back
on again.
• If the problem persists, inform seca Ser-
vice.
• Remove the load from the device.
• Switch off the device, then switch it back
on again.
• Switch off the device, then switch it back
on again.
• Switch off the device, then switch it back
on again.
• Ensure that the display cable is con-
nected correctly (
seca 797
• If the problem persists, inform seca Ser-
vice.
seca 103/452
Remedy
• Activate WiFi function
• Contact your administrator or hospital
technician
• Inform seca Service
• Ask the patient to step off the device and
wait until the Workflow LED goes out
• Check WiFi connection
• Deactivate
function
BMIF
• Repeat measurement procedure
• If the problem persists, contact your ad-
ministrator or hospital technician
• Contact your administrator or hospital
technician
• Cancel measurement procedure
• Repeat measurement procedure and ob-
serve proper sequence
• If the problem persists, contact your ad-
ministrator or hospital technician
• Contact your administrator or hospital
technician
• Inform seca Service
Activate WiFi function
).