Fitting the display head
24 • seca 787/797
4. Place one lock washer in each of the bores for the screws.
5. Tighten the screws using the T20 Allen key supplied (torque: 2 Nm).
6. Set the measuring station in an upright position.
Proceed as follows to fit the display head to the column:
1. Extend the upper telescopic element of the measuring rod approx. 10 cm.
2. Guide the cable(s) of the display head through the center channel of the col-
umn as far as the lower part of the device.
NOTE
One or two cables will be present, depending on the model:
– Connector cable (
– Connector cable and display cable (
3. Put the display head on the column.
4. Push the upper telescopic element into its lowest position.
5. Place one cross-head screw in each of the two side bores of the column.
6. Tighten the cross-head screws-head screws (torque: 1 Nm).
7. Place the measuring station on the floor with the display unit facing upward.
NOTICE!
Damage due to error in installation
If the cables are bent at too sharp an angle, they may be damaged
and thus cause a functional failure.
► Route all cables to prevent sharp bends and kinked connectors.