Activating/deactivating
switching of units of
measurement (UnIt) (variant-
Calibrating the measuring rod
40 • seca 787/797
With some device variants, you can switch the units of measurement using the
key. You can switch the function of the key on and off.
kg/lbs
dependent)
1. From the menu, select the
2. Confirm your selection.
ð The current setting is displayed.
3. Select the desired setting:
–
–
4. Confirm your selection.
ð The setting is adopted.
ð The menu closes.
Calibrate the measuring rod if the value on the display deviates from the value on
read-off edge 2.
(HCAL)
1. From the menu, select the
2. Confirm your selection.
ð
SEt
3. Set the measuring rod to 150 cm.
4. Confirm the setting.
ð The measuring rod is calibrated.
ð The menu closes.
UnIt
menu item.
On
OFF
HCAL
and
are displayed.
150.0 cm
menu item.