9. Technical Specification
Construction:
Adjustment:
Timer:
Power Supply:
Input Voltage:
Dimensions:
Weight:
Maximum Patient Load:
Maximum Additional
Equipment Load:
Product Features:
Compliance:
Environmental:
Ambient Temperature
(Operating):
Ambient Temperature
(Storage):
Relative Humidity:
Inspiration Healthcare Limited
W
inspiration-healthcare.com
Fabricated in mild steel, with powder coat paint finish, for durability and stability.
Compact, low-profile wheelbase to allow easy and convenient positioning, close to mother,
in all situations (natural birth, assisted delivery, caesarean section, etc)
Antistatic wheels with brakes
Electrically operated height adjustment with actuator for smooth adjustment and complete
stability
Ergonomic, unobtrusive platform to give easy access to infant
Patient platform height range from 70 cm to 110 cm from floor level
Foot pedal operated adjustment to leave hands free and facilitate aseptic techniques
Built-in timer with bright, backlit display to monitor cord clamping time
Audio alerts at 1, 2 & 3 minutes from delivery
Battery operated using 4×AA size 1.5 V alkaline cells (not supplied)
200 VA~ 50/60 Hz
230 V~ or 110 V~
Footprint 65 cm × 64 cm
Overall Height 79 cm (lowest position from floor level)
30 kg
10 kg
15 kg
Two Medirail mounting rails
Two mains power IEC outlets for Inspiration Healthcare approved accessories
The LifeStart has been tested and approved to comply with the following international
standards:
EN ISO 60601-1: 2006+A1:2013 (Edition 3.1) Medical electrical equipment: General
requirements for basic safety and essential performance
EN ISO 60601-1-2: 2007 Medical electrical equipment: General requirements for basic
safety and essential performance – collateral standard: Electromagnetic compatibility –
requirements and tests
The LifeStart is rated to a degree of Ingress Protection (IP) 30
15°C to 40°C (59°F to 104°F)
-10°C to 55°C (14°F to 131°F)
30% to 70%
16