Fig. 7: [START] button
Fig. 8: [STOP/OPEN] button
3.5.3
Controls
Fig. 9: [Rotary knob]
Fig. 10: [Mains switch]
Fig. 11: [Ramp-up and ramp-
down parameters] button
Fig. 12: [Cooling] button
Fig. 13: [PROG] button
AB5650en
■
The button flashes until the rotor is read in.
■
The button lights up during the centrifugation run for as long as the rotor
is not yet at a standstill.
■
The right side of the button lights up when the centrifuge is in ramp-
down. The rotor has not yet stopped.
■
The left side of the button lights up when the rotor is stationary.
■
The light on the left side of the button goes out when the lid is unlocked.
■
Setting the individual parameters.
Turning anticlockwise decreases the value.
Turning clockwise increases the value.
■
Switch the device on and off.
■
Ramp-up levels, parameters
Level 9 = shortest ramp-up time, Level 1 = longest ramp-up time.
■
Ramp-up time, parameters
Adjustable in 1 second increments.
■
Brake levels, parameters
1-9 = Linear braking curve
1b-9b = similar to an exponential braking curve
Level 9, 9b = shortest ramp-down time, ...Level 1, 1b = long ramp-
down time, Level 0 = unbraked ramp-down.
■
Ramp-down time, parameters
Adjustable in 1 second increments.
■
Brake cut-off speed, parameter N Brake
Adjustable from 50 RPM to the maximum rotor speed (N
ments of 10. Unbraked ramp-down takes place after reaching this
speed.
■
Start centrifugation run to pre-cool the rotor (only for devices with
cooling).
■
Centrifugation run, for pre-cooling of the rotor, takes place automatically
with the PREC program (PRECOOLING).
■
Retrieve programs and program links, parameter RCL (Recall).
Programs: Program locations 1 to 99. Program links: Program locations
A to Z.
■
Save programs and program links, parameter STO (Store).
99 programs can be stored (program locations 1 to 99).
Program location 0 serves as a buffer for the centrifugation data of
the last centrifugation run. No programs can be stored in this program
location.
25 program links can be stored (program locations A to Z, program
location J does not exist). A program link can consist of 20 programs.
Rev.: 24 / 01.2024
Device overview
), in incre-
max
19 / 73