Ambient conditions (EN / IEC 61010-1):
Installation site
Altitude
Ambient temperature
Humidity
Overvoltage category
(IEC 60364-4-443)
Pollution level
Device protection class
EMC:
Emitted EM interference,
EM interference immunity
Noise level
(rotor-dependent)
Dimensions:
Width
Depth
Altitude
Weight
Manufacturer
Model
Type
Mains voltage (±10%)
Mains frequency
power consumption
Power consumption
Refrigerant
max. capacity
max. permissible density
AB5650en
indoors only
up to 2000 m above sea level
2 °C to 35 °C
maximum relative humidity 80% for temperatures up to 31 °C,
decreasing linearly to 50% relative humidity at 40 °C.
II
2
I
not suitable for use in potentially explosive atmospheres.
EN / IEC 61326-1
Class B
£68 dB(A)
554 mm
706 mm
456 mm
approx. 101 kg
Andreas Hettich GmbH & Co. KG,
D-78532 Tuttlingen
ROTANTA 460 R
5660
5660-20
5660-50
5660-70
200-240 V 1~
50 Hz
1800 VA
8.5 A
R452A
4 x 1000 ml
1.2 kg/dm³
Rev.: 24 / 01.2024
Device overview
FCC Class B
715 mm
approx. 111 kg
5660-07
5660-77
200-240 V 1~
60 Hz
1900 VA
9.2 A
11 / 73