5.3
Setting up and connecting the centrifuge
Setting up the centrifuge
Connecting the centrifuge
AB5650en
WARNING
Risk of injury due to failing to maintain a sufficient distance to
the centrifuge.
As per EN / IEC 61010-2-020, no persons, hazardous
materials or objects may be present within a safety zone
of 300 mm around the centrifuge during a centrifugation
run.
A distance of 300 mm from the ventilation slots and ven-
tilation openings of the centrifuge must be maintained.
CAUTION
Risk of crushing and damage to the device due to it falling
down because of vibration-induced position alterations.
Place the device on a stable and level surface.
Select the installation surface dependent on the weight
of the device.
NOTICE
Damage to the samples and the device if the ambient
temperature exceeds or falls below the respective max-
imum/minimum permissible ambient temperature.
Comply with the maximum and minimum permissible
ambient temperatures for installation of the device.
Do not place the device next to a heat source.
Do not expose the device to direct sunlight.
Do not expose the device to frost.
Personnel:
Trained user
1.
Place the device on a stable and level surface.
2.
Maintain a distance of 300 mm around the device.
3.
Comply with the ambient conditions in the technical data (⮫ 
3.1 'Technical data' on page 10 ).
NOTICE
Damage to the device by unauthorised personnel
Tampering with and modifications to devices by unau-
thorised persons are at the operating organisation's own
risk and will result in the loss of all warranty and liability
claims.
Rev.: 24 / 01.2024