Unpacking after centrifugation
6.8
Centrifugation
6.8.1
Centrifugation in continuous operation
40 / 73
4.
If available, fold empty satellite bag(s) and pack differently depending
on the appropriate accessories and filling volume of the blood bag.
It is advantageous to fold the satellite bags and pack them on the
outside between the folded-down support plate and the outer wall of
the insert.
Ensure that the silicone plate does not slip when doing this.
If necessary, the silicone plate can be retained and held against the
loop when packing the satellite bag.
The position of the loop must be checked afterwards.
5.
Place the connections over the support plate so that the valves
cannot break.
Ensure that the hoses do not project out of the insert.
Stow hose sections projecting over the edge of the insert between the
folded-down support plate and the insert wall.
6.
If necessary, balance weights should be placed between the folded-
down support plate and the receptacle wall.
1.
Pull the satellite bag out of the insert while holding the silicone plate in
place with one hand.
2.
Slowly pull out the folded-down part of the support plate at the loop
provided for this purpose.
Return the support plate to its original shape in a controlled manner.
The folded-down part of the support plate may spring back and mix
blood components.
3.
Remove the remaining blood bag either together with the support
plate or after removing the support plate from the insert.
Personnel:
Trained user
1.
Set minutes, seconds and hours on
running program.
[START] button.
2.
Press the
�� The centrifugation run is started.
[START] button flashes until the rotor is read in.
The
The
[START] button lights up during the centrifugation run.
The timing starts at
The rotor speed or the resulting RCF value, the temperature in
the centrifuging chamber (only for centrifuges with cooling) and the
elapsed time are displayed during the centrifugation run.
Rev.: 24 / 01.2024
'0' or retrieve a continuously
'00:00' .
AB5650en