7.2
Programming
7.2.1
Write protection for programs
7.2.2
Opening or loading programs
7.2.3
Entering or changing programs
AB5650en
Write protection can be enabled or disabled when the rotor is at a standstill.
1.
Open the desired program.
[PROG] button.
2.
Press the
The RCL parameter is displayed.
3.
Press and hold the
The STO parameter is displayed.
'Set Protection = 1-' is displayed after 8 seconds.
[Rotary knob] to set '+' or '-' .
4.
Use
+ = The program is write-protected
- = The program is not write-protected
[START] button.
5.
Press the
The setting is stored.
1.
Press the
[PROG] button.
The RCL parameter is displayed.
[Rotary knob] to set the desired program location.
2.
Use the
[START] button.
3.
Press the
'Program recall...' is displayed briefly.
The centrifugation data of the desired program location is dis-
played
The previous data in the program location is overwritten
upon saving.
If "Protected !!" is displayed, the data in the program loca-
tion is write-protected and saving is not possible.
1.
Set the desired parameters.
[PROG] button repeatedly until the 'STO' parameter is dis-
2.
Press the
played.
[Rotary knob] to set the desired program location.
3.
Use the
If a "+" is displayed after the program location, the
data is write-protected.
The write protection must be removed before data
can be saved.
[START] button.
4.
Press the
Settings are stored in the desired program location.
'Program store...' is displayed briefly.
Rev.: 24 / 01.2024
[PROG] button.
Software operation