Device overview
Altitude
Weight
Manufacturer
Model
Type
Mains voltage (±10%)
Mains frequency
power consumption
Power consumption
Refrigerant
max. capacity
max. permissible density
max. speed (RPM)
max. acceleration (RCF)
max. kinetic energy
Obligation to perform
checks
(DGUV Rules 100-500)
(valid only in Germany)
Ambient conditions (EN / IEC 61010-1):
Installation site
Altitude
Ambient temperature
Humidity
Overvoltage category
(IEC 60364-4-443)
Pollution level
Device protection class
EMC:
14 / 73
456 mm
approx. 151 kg
Andreas Hettich GmbH & Co. KG,
D-78532 Tuttlingen
ROTANTA 460 RF
5675
5675-50
200-240 V 1~
50 Hz
1800 VA
8.5 A
R452A
4 x 1000 ml
1.2 kg/dm³
15000
24400
51000 Nm
yes
indoors only
up to 2000 m above sea level
5 °C to 35 °C
maximum relative humidity 80% for temperatures up to 31 °C,
decreasing linearly to 50% relative humidity at 40 °C.
II
2
I
not suitable for use in potentially explosive atmospheres.
Rev.: 24 / 01.2024
5675-01
5675-51
100-127 V 1~
60 Hz
max. 2000 VA
683 mm
approx. 140 kg
100 V 1~
50 Hz
AB5650en