Immunity
IEC 60601
Compliance
Test
Test Level
Level
Surge
±0.5kV,
±0.5kV,
±1 kV line(s)
±1 kV line(s)
IEC 61000-
to line(s)
to line(s)
4-5
±0.5kV, ±1kV,
±0.5kV, ±1kV,
±2 kV line(s)
±2 kV line(s)
to earth
to earth
Voltage
0 % U
; 0.5
0 % U
T
dips, short
cycle At 0°,
cycle At 0°,
interuptions,
45°, 90°,
45°, 90°,
and voltage
135°, 180°,
135°, 180°,
variations
225°, 270°
225°, 270°
on power
and 315°
and 315°
supply input
lines
0 % U
; 1
0 % U
IEC 61000-
T
cycle and
cycle and
4-11
70 % U
;
70 % U
T
25/30 cycles
25/30 cycles
Single phase:
Single phase:
at 0°
at 0°
0 % U
; 250
0 % U
T
cycles
cycles
This condition
causes the
RF controller
to shut
down and
then return
to standby
mode.
Power
30 A/m
30 A/m
Frequency
50 Hz
50 Hz
50/60Hz
Magnetic
Field
IEC 61000-4-8
Electromagnetic
Environment –
Guidance
Mains power quality
should be that of a
typical commercial or
hospital environment.
; 0.5
Mains power quality
T
should be that of a
typical commercial or
hospital environment.
If the user of the
NovaSure RF controller
requires continued
operation during power
mains interruptions, it is
recommended that the
; 1
T
NovaSure RF controller
be powered from an
;
uninterruptible power
T
supply or battery.
; 250
T
Power frequency
magnetic fields should
be that of a typical
commercial or hospital
environment.
Immunity
IEC 60601
Test
Test Level
Conducted
3 V
RF
0.15 MHz - 80
IEC 61000-
MHz
4-6
6 V in ISM and
amateur radio
bands between
0.15 MHz and
80 MHz
80% AM at 1 kHz
Radiated RF
3 V/m
IEC 61000-
80 MHz -
4-3
2.7GHz
80% AM at
1 kHz
27
Compliance
Electromagnetic
Level
Environment –
Guidance
Portable and mobile RF
communications equipment
should be used no closer to
any part of the RF generator,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended
separation distance
3 V
150 kHz to
0.15 MHz - 80
80 MHz
MHz
6 V in ISM and
amateur radio
bands between
0.15 MHz and
80 MHz
80% AM at 1 kHz
3 V/m
80 MHz to
800 MHz
80 MHz -
2.7GHz
800 MHz to
2.7 GHz
80% AM at
1 kHz
where P is the maximum
output power rating
of the transmitter in
watts (W) and d is the
recommended separation
distance in meters (m).
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
survey,
should be less
a
than the compliance level
in each frequency range.
Interference may
occur in the vicinity of
equipment marked with
the following symbol: