Télécharger Imprimer la page

BD Veritor Plus System Mode D'emploi page 20

Masquer les pouces Voir aussi pour Veritor Plus System:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 81
BD Veritor™ InfoWiFi Instructions
Proximity Fields From RF Wireless Communications Equipment
Test
a
Band
frequency
(MHz)
(MHz)
385
380 – 390
450
430 – 470
710
745
704 – 787
780
810
870
800 – 960
930
1720
1845
1700 –
1990
1970
2400 –
2450
2570
5240
5100 –
5500
5800
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting
antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance
is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while
it does not represent actual modulation, it would be worst case.
20
a
Service
Modulation
Pulse
TETRA 400
modulation
18 Hz
c
FM
GMRS 460,
± 5 kHz
FRS 460
deviation
1 kHz sine
Pulse
LTE Band
modulation
13, 17
217 Hz
GSM 800/900,
TETRA 800,
Pulse
iDEN 820,
modulation
CDMA 850,
18 Hz
LTE Band 5
GSM 1800;
CDMA 1900;
Pulse
GSM 1900;
modulation
DECT;
217 Hz
LTE Band 1, 3,
4, 25; UMTS
Bluetooth,
WLAN,
Pulse
802.11 b/g/n,
modulation
RFID 2450,
217 Hz
LTE Band 7
Pulse
WLAN
modulation
802.11 a/n
217 Hz
Maximum
Distance
b
power
(m)
(W)
b
1.8
0.3
2
0.3
b
0.2
0.3
b
2
0.3
b
2
0.3
b
2
0.3
b
0.2
0.3
TEST LEVEL
IMMUNITY
(V/m)
27
28
9
28
28
28
9

Publicité

loading