IMMUNITY
Phenomenon
Electrostatic Discharge
Radiated RF EM Fields
Proximity Fields From RF
Wireless Communications
Equipment
RATED Power Frequency
Magnetic Fields
Electrical Fast Transients /
Bursts
Surges Line-To-Line
Surges Line-To-Line
Conducted Disturbances
Induced By RF Fields
Voltage Dips
Voltage Interruptions
Radiated RF electromagnetic
field
Basic EMC Standard
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
AIM 7351731
Immunity Test Levels
± 4kV contact
± 2kV, ± 4kV, ± 8kV air
3 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
*See Table Below for Detailed
Frequency Ranges and Test
Levels
30 A/m
50/60 Hz
± 2 kV
100 kHz repetition frequency
± 0.5 kV, ± 1 kV
± 0.5 kV, ± 1 kV, ± 2 kV
3 V
0.15 – 80 MHz
6 V in ISM bands between
0.15 MHz and 80 MHz
80 % AM at 1 kHz
0 % UT; 0.5 cycle
AT 0°, 45°, 90°, 135°, 180°, 225°,
270° and 315°
0 % UT; 1 cycle
and
70 % UT; 25/30 cycles
Single phase: at 0°
0 % UT; 250/300 cycle
65 A/m @ 134.3 kHz
7.5 A/m @ 13.56 MHz
7.5 A/m @ 13.56 MHz
5 A/m, 12 A/m @ 13.56 MHz
3 V/m @ 433.92 MHz
54 V/m @ 860 – 690 MHz
54 V/m @ 2.45 GHz