Sterilization method and parameters
!
Use the implant system storage trays for sterilization and
sterile preparation.
Sterilization of the device may be accomplished by steam.
Aesculap does not recommend the device be sterilized by
"Flash" or chemical sterilization.
Surgical instruments may also be placed within an Aesculap
rigid sterilization container (sterile container) for processing
under generally accepted hospital in-use conditions.
The recommended sterilization parameters are as follows:
Sterilization
method
Pre-vacuum
44
Temp.
270—275 °F
4
S
Spinal System
Minimum exposure time
Wrapped
4 min
In a sterile
container
system
4 min