TROUBLESHOOTING

If an error occurs, the instrument display shows "Err" and the error number is also shown. The instru-
ment will now only react to the ENTER key. Pressing the ENTER key will attempt to restart the instru-
ment. Therefore, a reference run is automatically requested.
Problem
Instrument does
not react
Instrument does
not react
Instrument does
not react
Instrument does
not react
Tip drips/
instrument not sealed
or volume error
Display is dark
Error message
Possible cause
Battery discharged
Charge battery for at least 5
or faulty
min without operating, then
only operate with charging
cable attached until battery
is recharged. Replace bat-
tery if needed.
Faulty electronic
Send in the instrument
component
for repair.
Faulty electronic
Send in the instrument
component
for repair.
Unpredicted program
Confirm error by pressing
error
the ENTER key. The instru-
ment is reinitialized.
No battery inserted
Insert battery
Battery is defective
Replace battery
Faulty electronic
Send in the instrument
component
for repair.
Improper tip
Only use quality tips
Tip is not properly
Press tip in firmly/use
seated
other ejector-adjustment
clips.
Piston, nose cone or
Clean the instrument/
seal is contaminated
replace the seal.
or damaged
Oil piston.
Electrostatic dis-
Remove and insert
charge
the battery.
Faulty electronic
Send in the instrument
component
for repair.
Corrective action
ErgoOne E · User Manual · 31