ACCURACY TABLE
Final test values related to the nominal capacity (maximum volume)
indicated on the instrument and the indicated volume steps, obtained
when instrument and distilled water are equilibrated at ambient tem-
perature (20 °C/68 °F) and with smooth operation. According to DIN EN
ISO 8655.
Accuracy tolerances for the ErgoOne
Volume range
µl
0.5 -
10
2 -
20
10 - 200
50 - 1000
250 - 5000
* A = Accuracy, CV = Coefficient of Variation
E
®
Volume step
A*
µl
≤ ± %
10
1.0
5
1.5
1
5.0
20
1.0
10
1.5
2
5.0
200
0.8
100
1.2
20
4.0
1000
0.6
500
1.0
100
3.0
5000
0.6
2500
1.0
500
3.0
CV*
Increment
≤ %
µl
0.4
0.01
0.8
2.0
0.4
0.02
0.8
2.5
0.2
0.2
0.3
0.6
0.2
1.0
0.3
0.6
0.2
5.0
0.3
0.6
ErgoOne E · User Manual · 23
20 °C
Ex
Type of tip
µl
20
50
20
50
200
1000
5000