ELECTROPHORESIS (GEL) MODE
What to do
6.
Discharge liquid
in the GEL mode
7.
Initiate blow-out
8.
End pipetting
NOTE:
The GEL mode operates using a very slow discharge speed to prevent swirling of the samples. To
assure optimal discharging into a gel, this discharge speed is fixed for the GEL mode. This speed is
significantly slower than level 1 and cannot be selected individually.
How to do it
Press the pipetting key
once to discharge the
liquid. The rhombus is
shown in the display. The
liquid is discharged very
slowly.
Interrupt discharging
To interrupt discharging,
press the pipetting key
again. The display shows
the volume discharged
prior to interruption.
Press the ENTER key after
the last pipetting operation.
The display shows 'blo' for
blow-out.
Press the pipetting key
once to initiate the blow-
out process. The residual
liquid is discharged.
After the residual liquid is
discharged (blow-out), the
display moves back to the
start position.
Keys to press
Display readout
1x
1x
1x
1x
ErgoOne E · User Manual · 19