TROUBLESHOOTING

7.1 General errors

Symptom/message
Liquid drips out of
the pipette.
Aspiration capacity
reduced
No functional period
despite fully
charged battery.
Pipette loose.
Bubble formation in
the pipette during
liquid aspiration.
Rechargeable
battery will not
charge
28
· ErgoOne FAST · User Manual
Cause
Pipette adapter and/or
membrane filter
inserted incorrectly.
Pipette not inserted
far enough.
Pipette adapter
damaged.
Pipette damaged.
Seal damaged.
Liquid column
too high.
Speed too low.
Membrane filter
wetted.
Battery is too old.
Pipette adapter damaged.
Speed too high.
Rechargeable battery
is charged.
Remedy
Remove the pipette adapter
and membrane filter and
reinsert them.
Continue to carefully insert the
pipette.
Replace pipette adapter.
Replace pipette.
Replace seal.
Slowly press the aspirating button
down further.
Replace membrane filter.
Replace battery.
Replace pipette adapter
Apply less pressure to the aspirating
button.
Disconnect the Power supply.
Only recharge the battery if the
status display is blinking.