Masquer les pouces Voir aussi pour ErgoOne E:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

DISP MODE The Programs · Settings and Use
Program for discharging an aspirated liquid in pre-defined steps.
The volume aspirated will be a little bit more than actually needed.
Speed adjustment is described on page 9.
What to do
1.
Bring up the menu
2.
Select DISP
mode
3.
Confirm DISP
mode
4.
Prepare for
dispensing
5.
Set dispensing
step volume
6.
Confirm dispens-
ing step volume
20 · ErgoOne E · User Manual
How to do it
Press the MENU key three
times to bring ap the mode
selection menu.
'Mode' blinks.
Scroll through the modes
either the UP or DOWN key
until 'DISP' appears.
'Mode' continues to blink.
Press the ENTER key.
The Display now shows
'blo' for blow-out.
Press the pipetting key
once to move the piston
into the start position.
The arrow in the display
points upwards (aspira-
tion).
Press the arrow keys (+/-)
to set the volume. Holding
the arrow key down accel-
lerates the rate of change.
'VOL' blinks.
Press the ENTER key.
The display now shows the
new volume setting for the
dispensing steps and the
max. number of steps.
'Steps' blinks.
Keys to press
Display readout
3x
1x
1x
+
1x

Publicité

Table des Matières
loading