Maintenance
Microcentrifuge 24
6.5
Replacing fuses
The fuse holder is located under the mains power socket.
1. Disconnect the mains/power plug.
2. Open the fuse holder.
3. Replace faulty fuses and reinsert the fuse holder.
6.6
Decontamination before shipment
Please contact us before shipping a device.
If you are shipping the device to the authorized Technical Service for repairs or to your authorized dealer for disposal please note the following:
WARNING! Risk to health from contaminated device.
1. Observe the information on the decontamination certificate. It is available as a PDF document on our webpage
(www.starlab.click/decontamination).
2. Decontaminate all the parts you are going to dispatch.
3. Include the fully completed decontamination certificate in the shipment.
7
Troubleshooting
If you are unable to resolve the error with the suggested measures, please contact your local sales representative.
7.1
General errors
Problem
No display.
Centrifuge lid
cannot be opened.
Centrifuge cannot
be started.
Centrifuge shakes
when it starts up.
26
· Microcentrifuge 24 · Manual
Cause
No mains/power connection.
Mains/power outage.
The rotor is still running.
Mains/power outage.
The centrifuge lid is not closed.
The rotor is loaded asymmetrically.
Solution
Check the mains/power connection.
Check the fuse of the device.
Check the mains/power fuse of the lab.
Wait for the rotor to stop.
1. Check the fuse of the device.
2. Check the mains/power fuse of the lab.
3. Actuate the emergency lid release.
Close the centrifuge lid.
1. Stop the centrifuge and load the rotor
symmetrically.
2. Restart the centrifuge.