AUTOCLAVING
The pipette shaft of the ErgoOne
autoclaved at 121 °C (250 °F) at a pressure of 2 bar (30 psi) with
a holding time of at least 15 minutes according to DIN EN 285.
ATTENTION:
1.
Eject the pipette tip.
2.
Unscrew the pipette shaft from the grip.
3.
Autoclave the complete pipette shaft without
any further disassembling.
4.
Allow the pipette shaft to completely cool
and dry.
5.
Screw the pipette shaft into the grip again.
6.
Perform a reference run (rEF).
NOTE:
The effectiveness of the autoclaving must
be verified by the user.
Maximum reliability is obtained with
vacuum sterilization. We recommend the
use of sterilization bags.
If the pipette shaft is autoclaved frequently, then the
piston and seal should be greased with the supplied
silicone grease in order to preserve smooth movement.
REFERENCE RUN (REF)
A manual reference run must be completed each time the pipette shaft is reattached to
the handle. The reference run is needed to assure secure connection of the piston.
What to do
1.
Bring up rEF mode
2.
Perform the
reference run
After the reference run, the display automatically returns to the previous program.
NOTE:
26 · ErgoOne E · User Manual
E (highlighted in picture) can be
®
The handgrip can not be autoclaved!
How to do it
Simultaneously press the
MENU and the ENTER key
to activate the rEF mode.
Press the pipetting key
once to start the reference
run. A noise can be heard,
clearly indicating the func-
tion is being performed.
Keys to press
Display readout
1x
1x
S