EASY CALIBRATION
Revert to factory default settings
The continually displayed CAL symbol refers to a previously made
adjustment.
What to do
1.
Bring up the CAL
mode
2.
Deactivate CAL
mode
3.
Revert to factory
setting
IMPORTANT:
When the ErgoOne
performed, which means that the volume is changed
across the entire volume range of the pipette by the same
amount. It is recommend that the adjustment be per-
formed at 50% of the nominal volume.
NOTE:
The instrument is permanently adjusted for watery solu-
tions, but it can also be set for solutions with varying
density, viscosity and temperature. The ErgoOne
be adjusted in every mode, with the exception of the GEL
mode.
How to do it
Press and hold the MENU
key (> 3 sec) until CAL
appears. The display reads
'on'. 'CAL' blinks.
Press either the UP or
DOWN key to deactivate
the CAL mode. The display
changes from ‚on' to ‚off'.
'CAL' continues to blink.
Press the ENTER key.
The CAL symbol disap-
pears. The instrument has
now been reverted to fac-
tory default setting.
E is adjusted, a volume offset is
®
Keys to press
Display readout
1x
>3s
1x
E can
®