8.5.
Details of the process level
This level is active by default when the device is energized.
Value of the measured flow rate, displayed
in the unit chosen in the "UNIT" parameter of
12.6 l/s
the Configuration level.
Value of the main totalizer, total volume of
0......9
fluid counted by the device since the last
87654 l
reset.
Value of the daily totalizer (identified by a
dot after the volume units), volume of fluid
231 l.
counted by the device since the last reset.
+
0......9
> 2 s
Fig. 14: Details of the Process level
18
Resetting the daily
totalizer.
English
Type 8025/8035 battery powered
Operatingandcommissioning
8.6.
Details of the configuration level
To access the Configuration level, simultaneously press keys
ENTER
for at least 5 s.
This level comprises the following configurable parameters:
Choosing the display language
lANGUAGE
Choosing the flow rate unit, the
UNit
number of decimals and the unit
the totalizers are displayed in.
Entering the K factor of the fitting
K-FACtOR
used or have it defined through a
0......9
teach-in procedure.
Choosing the degree of attenuation
FiltER
of the measured flow rate varia-
tions, with effect on the displayed
flow rate.
Resetting both totalizers.
tOtAl
Going back to the Process level
END
and saving the settings.
Fig. 15: Diagram of the Configuration level