3.10 Dosing

Single component dosing
Functionality and operation are as described for "Control".
Multiple component dosing
Functionality and operation are as described for "Formulation".

3.11 Neutral measurement

To program the Function key, "Neutral measurement" must be selected in Setting
mode! For Function key settings, see Paragraph 1.13 or User mode 3720 (see
separate description 470.702.103) (see separate description 470.702.103) 
UCAL 4 Position 02 = Value 8.
The factor for entering the specific weight is the quantity of the measured
material relative to the calibrated unit weight (1 kg/1 g/1 lb).
Example: 1 kg paint = 0.823 litres (select UCAL1, Position 27 = 0.823)
Entering the conversion factor in Setting mode
Press and hold the Print button and then press the On/Off button for 5 seconds
until  appears.
Press the Print button to go into .
Using the Function key, scroll back to Position 26.
Use the Print button in the input field to go into the decimal places.
Set the number of decimal places with the Tare button and Function key.
Store with the Print button, the display goes to Position 27.
Use the Print button to go into the input field for the neutral measurement
factor.
Using the Tare button and Function key, set a number value in the flashing
space.
You can use the Zero button to move by one decade from right to left.
After setting the "Neutral Measurement" factor, store with the Print button,
Position 01 is displayed.
Press the Zero button,  appears on the display.
To save the setting and exit, press and hold the Print button then briefly press
the Zero button. The display returns to weighing mode.
www.soehnle-professional.com
60