These Safety Instructions contain our recommendations
for precautions in handling CERTOMAT® laboratory
shakers and incubation shaking cabinets from the
product program of the company Sartorius Stedim
Biotech GmbH, especially for

transport and installation

equipping and initial startup
operation
cleaning, maintenance, and service
These safety instructions are non-transferable and
meant exclusively for the equipment delivered. They are
not valid for other equipment. Additional, specific
safety regulations (i.e., legally or otherwise mandatory
rules) are not dealt with by these instructions. Please
consult your local authorities for further information.
Possible risks and dangers are marked with this symbol
and highlighted like this paragraph. Ignoring these
warnings may cause damage to the instrument or lead
to other material or personal damage.
Steps to be carried out with special care or special
aspects or issues to be considered are marked with
safety instructions formatted as this paragraph.
Identifies references to contents of these instructions or
[
..]
other documents. Labels of illustrations, sections, or
documents are in parentheses.
2
Transport and Installation
2.1
Transporting Equipment
The shakers and especially the incubation shaking
cabinets are heavy. Use only suitable transportation and
lifting tools when moving or setting up units.
Make sure clearances are sufficient along the transport
route and that the combined weight of lifting and
transportation tools does not exceed the maximum load
permitted for the floor or ground.
Use only qualified personnel to transport the equipment.
Make sure personnel is protected from injury during
transport.
Do not remove any transport protection before the
equipment has been transported to the site of
installation.
2.2
Requirements at the site of installation
1.
The floor or the laboratory bench where the equipment
is to be installed must be able to support the weight of
the fully equipped units and systems.
2.
The installation place should be sufficiently dimensioned
and feature a non-slip surface.
3.
The equipment has to be easily accessible for operation,
maintenance, and service tasks.
4.
The shakers and the incubation shaking cabinets may
transmit vibrations to the installation surface especially
when carrying a heavy load. For this reason, install the
shakers so that surrounding equipment is not affected.
5.
The power supply must match the rating indicated on
the type plate and must be grounded correctly.
Use a level to align the equipment after setting it up!
Before stacking equipment (max. 3 units), make sure
that the bottom most device is properly aligned
horizontally!