Activating/Deactivating Switching Of Units Of Measurement (Unit) (Variant-Dependent); Calibrating Measuring Rod (Hcal); Restoring Factory Settings (Reset) - Seca 787 Mode D'emploi Et Garantie

Masquer les pouces Voir aussi pour 787:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Activating/deactivating switching of
units of measurement (UnIt)
(variant-dependent)

Calibrating measuring rod (HCAL)

150,0 cm

Restoring factory settings (rESEt)

With some device variants, you can switch units of measurement using the
key. You can switch the function of the key on and off.
kg/lbs
1. Select the
2. Confirm your selection.
The current setting is displayed.
3. Select the desired setting:
4. Confirm your selection.
You exit the menu automatically.
If the value displayed in the display deviates from the value shown on read-off
edge 2, calibrate the measuring rod.
1. Select the
2. Confirm your selection.
and
SEt
3. Set the measuring rod to 150 cm.
4. Confirm the setting.
The measuring rod is calibrated.
You exit the menu automatically.
.
If you restore the factory settings (variant-dependent), the settings for the
following functions will be reset:
– Autohold (
– Permanent display of weight value (
– Filtering (
– Standby time (
– Switching units of measurement (
Units of height and weight are reset to metric units.
NOTE
seca 797
be retained.
1. Select the
2. Confirm your selection.
You exit the menu automatically.
item from the menu.
UnIt
On
OFF
item from the menu.
HCAL
are displayed.
150.0 cm
)
AHOLd
)
FIL
)
AOFF
only: If you restore the factory settings, the WiFi settings will
item from the menu.
rESEt
English
)
HOLd
)
UnIt
Operation • 35

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

797

Table des Matières