Calibration Setup - MSA ALTAIR 5X Manuel D'utilisation

Masquer les pouces Voir aussi pour ALTAIR 5X:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Calibration Setup

CALIBRATION OPTIONS
The Calibration Options menu has provisions to:
-
modify the calibration cylinder settings (CYLINDER SETUP)
-
enable/disable calibration due and to set the number of days (CAL DUE OPTIONS)
-
enable/disable the option to show the last cal date at turn on and (LAST CAL DATE)
When enabled, the date of the last device calibration displays during the turn-ON process.
-
enable/disable the option for password protected calibration (CAL PASSWORD)
When enabled, the device setup password must be entered prior to calibration.
Press:
-
the  button go to next page
the  button to go previous page
the q button to enter setup.
Setting Calibration Cylinder
This option has a dialog similar to the span calibration dialog.
The display shows all active sensors.
(1)
Press the q button to enter setup.
The screen for the first calibration cylinder displays.
(2)
Press
the  or  button to change the value.
the q button to confirm the setup.
With this confirmation the device automatically moves to the next cylinder setting.
(3)
Repeat the sequence for changing the required settings for all necessary gas values.
After the last setting is performed, the device returns to the Calibration Options menu.
ALTAIR 5X/ALTAIR 5X IR
Operation
US
32

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Altair 5x ir

Table des Matières