Zero Calibration

(1)
Press the  button for five seconds in Normal
Measurement page.
ZERO screen displays.
To skip the ZERO procedure and move directly to
the span calibration procedure, push the  button. If
no button is pushed for 30 seconds, the device
prompts user to perform a SPAN calibration before
device returns to the Normal Measurement page.
To perform ONLY a Fresh Air Setup at this time,
press the q button. The device then performs a
Fresh Air Setup as described in chapter 3.2. When
the Fresh Air Setup is complete, the device returns
to the normal Measure screen.
Press the  button to confirm the ZERO
(2)
screen, i. e. to execute zero calibration.
The message "SENSOR REFRESH"
displays, followed by the message "ZERO
CALIBRATION".
The "REFRESH" message does not
appear if a catalytic combustible sensor is
not installed.
ZERO calibration starts.
A progress bar shows the user how much
of the calibration has been completed.
During the first moments of a ZERO calibration, the
combustible sensor reading may be replaced by a
moving display of "PLEASE WAIT". This is normal.
After the ZERO calibration is completed
the device displays either
"ZERO CALIBRATION PASS"
or
"ZERO CALIBRATION FAIL".
Only if the device passes the zero calibra-
tion the SPAN screen displays.
ALTAIR 5X/ALTAIR 5X IR
Operation
US
42