Bump Test (BUMP page)

This page allows the user to perform an automated Bump Test on the device. To perform the test, the
(YES) button is pressed. See chapter 3.9 for details on performing the Bump Test.
If the  button is pressed, the Bump Test is not performed, and the display shows the next page in
the sequence (PEAK).
If the  button is pressed, the Bump Test is not performed, and the display reverts back to the normal
MEASURE page.
Peak Readings (PEAK page)
Monochromatic display
PEAK
This page shows the highest levels of gas recorded by the device since turn-ON or since peak readings
were reset.
To reset the peak readings:
(1)
Access the PEAK page.
(2)
Press the  button.
This page can be de-activated through MSA Link software.
Minimum Readings (MIN page)
Monochromatic display
MIN
This page shows the lowest level of oxygen recorded by the device since turn-ON or since the MIN
reading was reset. It is only shown if an oxygen sensor is installed and enabled.
To reset the MIN reading:
(1)
Access the MIN page.
(2)
Press the  button.
ALTAIR 5X/ALTAIR 5X IR
Description
Color display
Color display
US
20